|
This case was reviewed and investigated according to the manufacturer¿s policy.The pioneer plus catheter was returned for evaluation.Visual inspection found a bent scanner.During functional testing, the catheter was plugged into a laboratory intrasight system, was recognized; however, failed to produce an image.Further tests were performed, but also failed the wire bond, apt, and image tests.Block h6: the probable cause of the bent scanner was likely damaged during use/handling.Strain, impact, and forces associated with use can affect the integrity of the electrical connections within the device.Blocks h7, h9 & h10: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|