• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RDCE FRCPS W/ PTS BRD; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RDCE FRCPS W/ PTS BRD; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Catalog Number 71173377
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/16/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during an internal fixation surgery, the tip of the rdce frcps w/ pts brd fractured while dr.Tried to compress fracture by grabbing plate and screw.Tip of the forceps broke off; piece did not fall inside the wound.Broken piece landed outside the patient on the floor.The procedure was resumed, without any delay, using a s+n back-up device.No injury was reported as a consequence of this issue.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.However, the photographs were reviewed, and revealed that the tip of the forceps is broken.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed a similar event for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RDCE FRCPS W/ PTS BRD
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key21318194
Report Number1020279-2025-00233
Device Sequence Number18109311
Product Code JDG
UDI-Device Identifier03596010519535
UDI-Public3596010519535
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71173377
Device Lot Number21ATCD001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/16/2025
Supplement Date Received by Manufacturer03/05/2025
Initial Report FDA Received Date02/05/2025
Supplement Report FDA Received Date03/10/2025
Date Device Manufactured03/11/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-