| Model Number |
C2PIVL4012 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Myocardial Infarction (1969)
|
| Date of Event |
04/22/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device referenced in this report was not returned to shockwave medical, inc.Therefore, a physical examination cannot be performed.The cause of the myocardial infarction could not be definitively determined based on the information available.There was no ivl malfunction reported.Shockwave medical has controls in place to ensure devices are built to approved procedures and meet lot release acceptance criteria prior to being distributed.A review of the manufacturing and test documentation for subject lot does not reveal any issues with the manufacturing of the device.The device passed all of shockwave medical, inc.Acceptance criteria prior to shipping.
|
| |
|
Event or Problem Description
|
|
This event was reported to shockwave via the post market empower cad clinical study.A shockwave c2+ coronary intravascular lithotripsy (ivl) catheter was used to treat a lesion in the mid left anterior descending artery.120 pulses were successfully delivered at 4 atm.There was no ivl malfunction reported.This event was sent to the clinical events committee (cec) for adjudication of the myocardial infarction (mi).The cec determined that the mi was a non q-wave mi attributable to the target vessel.They determined that the mi was possibly related to the study device and definitely related to the procedure.Additionally, it was reported that the septal was closed and there was heart block, but no q wave.
|
| |
|