• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHOCKWAVE MEDICAL, INC. SHOCKWAVE C2 CORONARY INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHOCKWAVE MEDICAL, INC. SHOCKWAVE C2 CORONARY INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER Back to Search Results
Model Number C2PIVL4012
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Myocardial Infarction (1969)
Date of Event 04/22/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device referenced in this report was not returned to shockwave medical, inc.Therefore, a physical examination cannot be performed.The cause of the myocardial infarction could not be definitively determined based on the information available.There was no ivl malfunction reported.Shockwave medical has controls in place to ensure devices are built to approved procedures and meet lot release acceptance criteria prior to being distributed.A review of the manufacturing and test documentation for subject lot does not reveal any issues with the manufacturing of the device.The device passed all of shockwave medical, inc.Acceptance criteria prior to shipping.
 
Event or Problem Description
This event was reported to shockwave via the post market empower cad clinical study.A shockwave c2+ coronary intravascular lithotripsy (ivl) catheter was used to treat a lesion in the mid left anterior descending artery.120 pulses were successfully delivered at 4 atm.There was no ivl malfunction reported.This event was sent to the clinical events committee (cec) for adjudication of the myocardial infarction (mi).The cec determined that the mi was a non q-wave mi attributable to the target vessel.They determined that the mi was possibly related to the study device and definitely related to the procedure.Additionally, it was reported that the septal was closed and there was heart block, but no q wave.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKWAVE C2 CORONARY INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER
Common Device Name
SHOCKWAVE C2 CORONARY INTRAVASCULAR LITHOTRIPSY (IVL) CATHETER
Manufacturer (Section D)
SHOCKWAVE MEDICAL, INC.
5403 betsy ross drive
santa clara 95054
Manufacturer (Section G)
SHOCKWAVE MEDICAL, INC
5403 besty ross drive
santa clara CA 95054
Manufacturer Contact
alexis weil
5403 betsy ross drive
santa clara 95054
MDR Report Key21318442
Report Number3015053858-2025-00014
Device Sequence Number15420266
Product Code QMG
UDI-Device Identifier00195451000010
UDI-Public(01)00195451000010(17)251031(10)01A231017A
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P200039/S008
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model NumberC2PIVL4012
Device Catalogue NumberC2PIVL4012
Device Lot Number01A231017A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2025
Initial Report FDA Received Date02/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
POST-DILATATION CATHETER - 2.75 X 15 - ABBOTT NC; POST-DILATATION CATHETER - 3.00 X 15 - ABBOTT NC; POST-STENT DILATATION CATHETER - 2.75 X 15 - UNK; POST-STENT DILATATION CATHETER - 3.00 X 15 - UNK; POST-STENT DILATATION CATHETER - 3.50 X 8 - UNK; PRE-DILATATION CATHETER - 2.50 X 15 - ABBOTT NC
Outcome Attributed to Adverse Event Other;
Patient Age77 YR
Patient SexFemale
Patient EthnicityHispanic
-
-