|
The patient's daughter, requested assistance with a clogged luer port on an express mini 500 cd.The patient has been using the drain at home for the past five days.I reviewed the off-label use of the product in a home setting.She reported that the cd currently contains approximately 200cc of fluid, which she has been draining once daily via the luer port.Advised her to contact the patient¿s care team or take the patient to the hospital er for a cd replacement.
|
|
Investigation summary: complaint involves use of an express mini 500 (p/n 16400) in an outpatient setting, in which the sampling port reportedly became clogged due to attempts to empty the fluid collected.In these cases, the drain is set up on a patient by a clinician and then that patient is sent home with the drain where they are then responsible for its use and interpretation.This puts the responsibility of using the drain on a person who is not necessarily a trained medical professional with any knowledge of how to use and interpret the device.This leads to the several types of complaints outlined below, all of which share a root-cause of user error due to the user of the drain being unfamiliar with proper use of the drain.Sampling port clogged: this typically occurs when the user attempts to empty the drain with a syringe through the sampling (luer) port in order to extend the use of the drain.This can lead to collected fluid becoming trapped in the sampling port and solidifying, preventing future use of the port.This port is intended for clinicians to take samples of the drain fluid, not for routinely removing large amounts of fluid in order to extend the use of the drain.The ifu instructs the user to replace the drain when it becomes full.It also states that the use of the luer port is intended only for sampling patient drainage and that the user should not use the port or any other means to attempt to empty fluid from the collection chamber.In addition to the instructions described for each type of complaint, the ifu states that "the express mini 500 is restricted for use in a healthcare facility.The express mini 500 should not be used for outpatient drainage." this type of complaint has been thoroughly investigated and is known to be cause by user error so further dhr review is not required to understand the issue.The ifu provides adequate instructions for the setup and use of the drain, as well as instructions about the intended users and environments.Outpatient use complaints of express mini 500 devices are confirmed.Device nonconformances of outpatient issues are not confirmed.The root cause of outpatient use complaints of express mini 500 complaints is user error.
|