MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW CVC SET: 2-LUMEN 7 FR X 20 CM; CATHETER INTRAVASCULAR THERAPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW CVC SET: 2-LUMEN 7 FR X 20 CM; CATHETER INTRAVASCULAR THERAPE Back to Search Results
Catalog Number CV-47702-E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "the doctor found the ars leak during used on the patient".They changed a new one to resolve the issue.No harm for the patient.The patient condition is fine.No medical intervention was required.
 
Event or Problem Description
It was reported that "the doctor found the ars leak during used on the patient".They changed a new one to resolve the issue.No harm for the patient.The patient condition is fine.No medical intervention was required.
 
Additional Manufacturer Narrative
(b)(4).The customer returned one arrow raulerson syringe (ars), introducer needle, dilator, transduction probe, guidewire assembly, and clamps for analysis.Signs of use were observed on some of the returned components, including the ars and introducer needle.Initial visual analysis of the ars revealed no obvious defects or anomalies.After failing functional testing, a hole was observed in the valves when viewing the ars valve through the handle.The ars was not able to draw and aspirate sufficient water with or without the introducer needle.Small amounts of water could be drawn but the sample did not function as intended.The module requirement document for raulerson syringes (amrq-000113 rev03) was reviewed to determine requirements for air/water leakage.The document notes a deviation from iso 7886-1: "the freedom of air and liquid leakage past the piston requirement is design restrictive and is intended for an injection-intended syringe, not the ars.The opening in the center of the piston that allows passage of the inner cannula prohibits the ars from meeting the pressure and vacuum requirements as dictated by the standard.However, because the intended use of the ars is to allow aspiration of blood to ensure venous placement of the introducer needle and to aid in the insertion of the spring wire guide, the leakage requirements of a standard syringe are not applicable to the ars." a vacuum test was performed on the ars syringe in order to verify that the internal valves within the plunger body were intact.With the plunger body at the bottom of the syringe, the tip of the ars was occluded, and the plunger was pulled back until it stopped.With the tip of the ars still occluded, the plunger was released and did not snap back into a position = 1cc from the starting position.Therefore, the internal valves of the ars were not functioning as intended.A device history record review was performed and no relevant findings were identified.The ifu provided with the kit informs the user, "insert introducer needle or catheter/needle with attached syringe or arrow raulerson syringe (where provided) into vein and aspirate." the customer report of an ars leak was confirmed through investigation of the returned sample.During functional testing, the ars was not able to properly draw water and did not pass additional testing performed.Therefore, based on the customer report and the sample received manufacturing likely caused or contributed to this event.A non-conformance was initiated to further investigate this issue.Teleflex will continue to monitor and trend on reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 2-LUMEN 7 FR X 20 CM
Common Device Name
CATHETER INTRAVASCULAR THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key21321239
Report Number3006425876-2025-00130
Device Sequence Number18266259
Product Code FOZ
UDI-Device Identifier10801902102536
UDI-Public(01)10801902102536(11)220922(17)240829(10)71F22J1153(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K993691
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date08/29/2024
Device Catalogue NumberCV-47702-E
Device Lot Number71F22J1153
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2025
Supplement Date Received by Manufacturer03/21/2025
Initial Report FDA Received Date02/06/2025
Supplement Report FDA Received Date04/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-