Model Number MAX-N-I
Medical Device Problem Code Failure to Analyze Signal (1539)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that there was a delay in care for neonatal patient due to a reading issue with the sensor, which resulted in no spo2 reading.The sensor light appeared to be less bright.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 correction: h8.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that there was a delay in care for neonatal patient due to a reading issue with the sensor, which resulted in no spo2 reading.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Common Device Name
OXIMETER
Manufacturer (Section D)
NELLCOR PURITAN BENNETT MEXICO SA
boulevard insurgentes 19030
tijuana 22225
MX  22225
Manufacturer (Section G)
NELLCOR PURITAN BENNETT MEXICO SA
boulevard insurgentes 19030
tijuana 22225
MX   22225
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21321415
Report Number1282497-2025-00060
Device Sequence Number7918497
Product Code DQA
UDI-Device Identifier20884522040379
UDI-Public20884522040379
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K012891
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAX-N-I
Device Catalogue NumberMAX-N-I
Device Lot Number242810090H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2025
Supplement Date Received by Manufacturer04/07/2025
Initial Report FDA Received Date02/06/2025
Supplement Report FDA Received Date04/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-