• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH AIM-ARM RADIOLUC F/GREATER TROCHANT FEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH AIM-ARM RADIOLUC F/GREATER TROCHANT FEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 03.033.003
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2025, the patient underwent orif surgery with the products in question.In the surgery, the guide pin was inserted using the products in question, but it did not pass through the intramedullary nail and slipped forward and could not be inserted in the proper direction.The guide pin was inserted by tilting the aiming arm backward, but it slipped forward again and could not be inserted.The arm was repositioned, but it was led toward the guide pin that was already open and could not be inserted.Since the guide pin was not inserted in the correct direction, we changed the screw position to the position of the downstroke and transverse instead of fixing the recontouring mode.The reasons for the failure were that the intramedullary nail was placed in a patient with a tight medullary cavity, the guide pin was guided in the direction of the guide pin that was placed first, the position of the aiming arm was not changed much because the intramedullary nail was placed tightly, and the sleeve may have been slightly deflected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIM-ARM RADIOLUC F/GREATER TROCHANT FEM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key21322100
Report Number8030965-2025-01353
Device Sequence Number7866033
Product Code HSB
UDI-Device Identifier07612334121790
UDI-Public(01)07612334121790
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K172157
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.033.003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/26/2025
Supplement Date Received by Manufacturer02/20/2025
Initial Report FDA Received Date02/06/2025
Supplement Report FDA Received Date02/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-