SYNTHES GMBH 2.7MM TI VA LCP CLAVICLE PLATE SHAFT/ CS1/ LEFT/ STERILE; PLATE, FIXATION, BONE
|
Back to Search Results |
|
| Catalog Number |
04.112.620S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/28/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, the patient underwent a surgery with the plate in question.The surgery was completed successfully with no surgical delay.After the surgery on (b)(6) 2025, the sales rep was informed that the plate dislocated.When the sales representative went to the hospital to confirm the x-ray, it was confirmed that the screw remained in the plate and the distal bony fragment, which was lumped with the shoulder joint, was displaced caudally by the weight of the arm.No screw fracture damage was seen.The revision surgery is scheduled on (b)(6) 2025, to remove the plate and replace with other company¿s implants.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.Investigation summary: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: part# 04.112.620s, lot # 27937p0, manufacturing site: werk selzach logistik, manufacturing date: 16.July.2024, expiration date: (b)(6) 2034, supplier: (b)(4).A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|