|
D10 concomitant products: 88863306-56, 88863306-56 2-0, 6x90 cm cv305 da7p, (lot # d2l1052y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
According to the reporter, intra-operatively, the needles of two sutures detached while preparing suture to suture mitral valve during a mitral valve replacement procedure.No device component fell into the patient¿s cavity.Another device was used to complete the case.There was no patient injury.
|