• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE SG; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE SG; IMPLANTABLE LEAD Back to Search Results
Model Number 0283
Medical Device Problem Codes Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the right ventricular (rv) lead exhibited high, out-of-range pace impedance measurements of greater than 3000 ohms.Technical services (ts) was contacted, and ts discussed options.The rv lead remains in service.No adverse patient effects were reported.
 
Event or Problem Description
It was reported that the right ventricular (rv) lead exhibited high, out-of-range pace impedance measurements of greater than 3000 ohms.Technical services (ts) was contacted, and ts discussed options.The rv lead remains in service.No adverse patient effects were reported.Additional information was received indicating rv lead noise and out-of-range pace impedances resulted in signal artifact monitor (sam) disabling minute ventilation (mv).The rv lead remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE SG
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key21327334
Report Number2124215-2025-06568
Device Sequence Number15469556
Product Code NVY
UDI-Device Identifier00802526431388
UDI-Public00802526431388
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/27/2018
Device Model Number0283
Device Catalogue Number0283
Device Lot Number328801
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2025
Supplement Date Received by Manufacturer03/27/2026
05/21/2026
Initial Report FDA Received Date02/06/2025
Supplement Report FDA Received Date03/30/2026
05/27/2026
Date Device Manufactured02/27/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
-
-