| Catalog Number |
806316 |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the critical care nurses reported in an online survey that the all-silicone foley catheter, 2-way, standard length, 30 cc balloon, 16 fr was not successfully placed in the patient because of the size was rather big.They swapped to pig tail size 8 and then connected to size 16 ch tubing.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.No actions can be taken at this time since a root cause was not identified.A dhr could not be performed since no lot number was provided.The instructions for use were found adequate and state the following: warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage silicone and may cause the balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local state, and federal laws and regulations.Cautions: federal (u.S.A) law restricts this device to sale by or on the order of a physician.Cautions: do not aspirate urine through drainage funnel wall.Storage: store catheter at room temperature away from direct exposure to light, preferably in the original box.Note: aggressive traction, particularly in the presence of suturing, is not recommended for 100 percentage silicone foley catheters.Sterile unless package is opened or damaged.Single patient use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip syringe.Do not use needle.Catheters should be replaced in accordance with the cdc guideline guideline for prevention of catheter-associated urinary tract infection.At the onset or first signs of a urinary tract infection, catheter encrustation, or any other catheter-related adverse effect, the catheter should be replaced.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe stick in the valve.If you notice slow or no deflation, re- seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Recommended inflation capacities.3cc balloon: use 5ml sterile water 5cc balloon: use 10cc sterile water 30cc balloon: use 35cc sterile water do not exceed recommended capacities.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the critical care nurses reported in an online survey that the all-silicone foley catheter, 2-way, standard length, 30cc balloon, 16 fr was not successfully placed in the patient because of the size was rather big.They swapped to pig tail size 8 and then connected to size 16ch tubing.Per additional information received on 08mar2025, it was reported that the complaint was not due to device related failure, but this occurred due to the complexity of their care needs.There was no negative impact to the patient as a result of the device.Pig tail catheter was considered due to presence of bladder tumor and patient experiencing significant hydronephrosis but was unfit for surgery at the time.As urethral catheterization failed due to bladder outlet obstruction.Pigtail catheter was then used.
|
| |
|