The reported event could be confirmed since the ct scan shows the clear loosening and subsidence of the tibial component.It is dislocated proximally and a periprosthetic fracture at the anterior tibial aspect if the result of the loosening and migration.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.Upon further investigation of the ct scans, healthcare professionals opined that - the ct scan shows the clear loosening and subsidence of the tibial component.It is dislocated proximally and a periprosthetic fracture at the anterior tibial aspect is the result of the loosening and migration.There is some radiolucence visible at the talar component, but no clear loosening or migration can be confirmed.The underlying cause of the loosening is likely to be patient related due to poor bone substance.However, failure of total ankle replacement tar over time is a known complication.In case of a planned revision procedure a medical opinion aiming to determine the root cause of the failure is part of the internal investigation process.Sufficient radiological and clinical information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, the assessment is limited.No conclusive statement can be provided, because key clinical information e.G.Clinical status, exact symptoms and range of motion of the patient, assessment of the treating physician is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this decisive evidence.Due to these limitations, we are unable in such cases to provide statements about the patient, the procedure and or the device in relation to cause of the failure.Based on investigation the issue is caused most likely due to patient related factor i.E.Poor bone substance but more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
|