• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BIO-COMP, SWIVELOCK C, FT, 4.75X19.1MM; BIO SOFT TISSUE FIXATN FASTNR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BIO-COMP, SWIVELOCK C, FT, 4.75X19.1MM; BIO SOFT TISSUE FIXATN FASTNR Back to Search Results
Model Number BIO-COMP, SWIVELOCK C, FT, 4.75X19.1MM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a speed bridge rotator cuff repair surgery the anchor broke off while it was screwed in.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.Update dw 28-jan-2025: further information was provided that the broken tip was retrieved out of the patient.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used during insertion, improper bone prep and/or prying/leveraging the device during use.The bone quality of the patient and the information of the instrument used to prepare the bone socket for insertion of the implant were not provided.The dfu for this device, dfu-0087-eo revision 4, gives the following warning: 6.Bioabsorbable only: attempting implantation into hard, dense bone and/or drilling/punching smaller diameter holes than recommended may cause failure (breakage) of the implant during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-COMP, SWIVELOCK C, FT, 4.75X19.1MM
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21333830
Report Number1220246-2025-00300
Device Sequence Number3848921
Product Code MAI
UDI-Device Identifier00888867098022
UDI-Public(01)00888867098022(11)231114(17)271031(10)15167637
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K101823
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBIO-COMP, SWIVELOCK C, FT, 4.75X19.1MM
Device Catalogue NumberAR-2324BCCT-1
Device Lot Number15167637
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2025
Supplement Date Received by Manufacturer10/20/2025
Initial Report FDA Received Date02/07/2025
Supplement Report FDA Received Date10/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/14/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-