MAUDE Adverse Event Report: MICROLINE SURGICAL INC RENEW SUPER ATRAUMATIC GRABBER TIP, DISPOSABLE; RIGID ENDOSCOPIC TISSUE MANIPULATION FORCEPS, SINGLE-USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROLINE SURGICAL INC RENEW SUPER ATRAUMATIC GRABBER TIP, DISPOSABLE; RIGID ENDOSCOPIC TISSUE MANIPULATION FORCEPS, SINGLE-USE Back to Search Results
Model Number 3622
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2024
Type of Reportable Event Malfunction
Event or Problem Description
After surgery, it was the sterile staff who discovered that a rubber part from a grabber tip was on another instrument.The rubber probably came loose when they took out the forceps with the grabber tip and when they drove down a laparoscopic needle driver (in the trocar), the rubber piece had gotten stuck there.No damage was done and the surgery was not delayed.
 
Additional Manufacturer Narrative
At this time, microline surgical is awaiting the suspected device back so an investigation can be conducted.Once the results of the investigation are available, a final adverse event report will be submitted.
 
Additional Manufacturer Narrative
An investigation of the returned product was able to confirm the reported issue of part of the gabber tip coming loose.The scissor tip was received, decontaminated, and visually inspected.The heat shrink was no longer attached to the grabber tip.It is believed that this is a result of the grabber tip not being subjected to heat for the appropriate amount of time.Microline surgical inc.Has received similar complaints recently and as a result initiated a corrective action (car-0200) in (b)(6) 2024 to address this problem.Immediate containment carried out by microline surgical inc.Included inspecting all tips in process and in our stockroom made on this assembly line for this defect.Additionally, our quality assurance team performed an independent verification of heat shrink prior to release for every lot of tips being manufactured on the affected assembly line until the car investigation was complete.The root cause identified through the investigation of car-0200 was that the tips were not processed in accordance with the work instruction.The parts were not subjected to elevated temperature in an oven as the assembly process calls for.Therefore, the heat shrink did not firmly attach to the tip assemblies.When investigating further the following factors were identified as contributing to the process failure: -the state of work of material was not properly identified during break times and the end of the day.-the work area had no visible staging area to identify status of parts (awaiting oven versus already processed through oven).- insufficient inspection and training of operators.-the oven process log does not include the number of tips processed.There is no method to reconcile if all parts were placed in the oven.For the implementation stage of car-0200 the assembly line work area was rearranged to provide a separate holding area for tips processed through the heat shrink oven.Product is held in this area until it is 100% inspected and released for further processing.Our assembly work instruction was updated to include more detail on how to perform a tactile pull/push test which will ensure the heat shrink is secure.Our inspection procedure was revised to include a step on visually examining the shrunk tubing.A step was also added to our packaging work instruction as another level of verifying that the heat shrink is firmly attached.Lastly, the oven process log used in tip assembly was revised to record the quantity of tips processing through the oven.All operators have been trained to the updated procedures.Car-0200 is currently in the verification stage where we will ensure the improvements implemented are still being practiced and we will monitor customer complaints for this defect.
 
Event or Problem Description
After surgery, it was the sterile staff who discovered that a rubber part from a grabber tip was on another instrument.The rubber probably came loose when they took out the forceps with the grabber tip and when they drove down a laparoscopic needle driver (in the trocar), the rubber piece had gotten stuck there.No damage was done and the surgery was not delayed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENEW SUPER ATRAUMATIC GRABBER TIP, DISPOSABLE
Common Device Name
RIGID ENDOSCOPIC TISSUE MANIPULATION FORCEPS, SINGLE-USE
Manufacturer (Section D)
MICROLINE SURGICAL INC
50 dunham rd suite 1500
beverly MA 01915
Manufacturer (Section G)
MICROLINE SURGICAL INC
50 dunham rd suite 1500
beverly MA 01915
Manufacturer Contact
scott marchand davis
50 dunham rd suite 1500
beverly, MA 01915
9789229810
MDR Report Key21334129
Report Number1223422-2025-00002
Device Sequence Number3849430
Product Code GEI
UDI-Device Identifier00811099010142
UDI-Public00811099010142
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K213127
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3622
Device Catalogue Number3622
Device Lot Number00173425
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2025
Supplement Date Received by Manufacturer02/07/2025
Initial Report FDA Received Date02/07/2025
Supplement Report FDA Received Date05/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-