Model Number G154
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was exhibiting pacing impedance high out of range and continues to rise.The device remains in service.No additional adverse patient effects were reported.
 
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was exhibiting pacing impedance high out of range and continues to rise.The device remains in service.No additional adverse patient effects were reported.Additional information was obtained; the remote communicator displayed a red call doctor icon due to pacing lead impedance out of rang.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN? CRT-D
Common Device Name
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key21334313
Report Number2124215-2025-06782
Device Sequence Number8549872
Product Code NIK
UDI-Device Identifier00802526534652
UDI-Public00802526534652
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/09/2020
Device Model NumberG154
Device Catalogue NumberG154
Device Lot Number103395
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2025
Supplement Date Received by Manufacturer08/08/2025
Initial Report FDA Received Date02/07/2025
Supplement Report FDA Received Date09/02/2025
Date Device Manufactured08/24/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
-
-