| Model Number |
KD-V411M-0730 |
| Medical Device Problem Codes |
Break (1069); Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: the coated portion of the cutting wire was torn.Based on the results of the investigation, it was determined that the cutting wire was broken at proximal end site.A definitive root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor the field performance of this device.
|
| |
|
Event or Problem Description
|
|
It was reported that while using the single use 3-lumen sphincterotome v the cutting wire of the papillotome broke or was damaged when cutting under tension.The reported malfunction occurred during a therapeutic endoscopic retrograde cholangio-pancreatography.The procedure was completed using a similar device.There were no reports of patient harm.
|
| |
|