Model Number VICMO12.6
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cataract (1766); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/15/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information: b5 - the reporter indicated that capsular tear was reported in error.Lens opacity was noted to be found along with the low vaulting during a check up.Claim#: (b)(4).
 
Event or Problem Description
A facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient experienced low vaulting, with no lens rotation, and capsular tear.The lens was exchanged with a longer length lens and the problem was resolved.The cause of the event was reported as unknown.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
 
Additional Manufacturer Narrative
H6 - health effect clinical code: 4581 - capsular tear.Manufacturer narrative: secondary surgical intervention to remove/replace/reposition the lens, is identified in the labeling as a known potential adverse event following icl implantation.Claim#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key21339512
Report Number2023826-2025-00236
Device Sequence Number8598535
Product Code MTA
UDI-Device Identifier00840311302109
UDI-Public(01)00840311302109(11)230428(17)260331
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Model NumberVICMO12.6
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2025
Supplement Date Received by Manufacturer03/05/2025
Initial Report FDA Received Date02/07/2025
Supplement Report FDA Received Date03/06/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR