This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: e3: reporter is a j&j sales representative.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee synovectomy procedure for knee synovitis it was observed that the vapr3 footswitch did not operate for coagulation.It was reported that the coagulation did not work.It was reported that the device connection was plugged and unplugged to the main unit, but the situation did not change.There were no adverse consequences to the patient, or surgical delay.No additional information could be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: correction h4: the device manufacture date was reported as may 2, 2019 on the initial report.Please note that the date of manufacture has been updated accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: investigation summary the complaint device was not returned to j&j medtech orthopaedics, therefore unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If the complaint device is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|