C.R. BARD, INC. (BASD) -3006260740 M.R.I. IMPLANTABLE PORT, HICKMAN SINGLE-LUMEN, 9.6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Catalog Number |
0602660 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562); Deformation Due to Compressive Stress (2889); Migration (4003)
|
| Health Effect - Clinical Codes |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/17/2003
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported through litigation process that sometime later port placement procedure, the catheter fractured.However, the current status of the patient is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported through litigation process that sometime later port placement procedure, the catheter fractured.However, the current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No medical records were provided for review.Therefore, the investigation is inconclusive for the reported catheter fracture as no objective evidence was provided for review.Furthermore, the clinical conditions alleged in the complaint cannot be confirmed.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date), g3, h6 (method, component).H11: d4 (unique identifier (udi)# h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported through litigation process that, on (b)(6) 2002, a patient underwent port placement in the chest via the right subclavian vein.Approximately, nine months and fourteen days later port placement on (b)(6)2003, the catheter had allegedly fractured, and a fragment migrated to the atrium.Subsequently, the port and catheter were removed on the same day.It was further reported that tubing on the end of the device was crimped.However, the current status of the patient remains unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional information was identified in the medical record.Based on the identified information the file was reassessed for reportability and determined to be reportable as serious injury.Manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: although the physical device was not returned for evaluation, medical record was provided for review.Medical record alleges that, patient was implanted with power port isp m.R.I.Implantable port for chemotherapy of lymphoma via right subclavian vein.Fluoroscopic imaging confirmed the catheter tip at the svc right atrial junction with no kinks.The port returned blood easily and was flushed with heparinized saline.Approximately nine months and fourteen days later, it was found that a portion of the port broke and had to be retrieved from the atrium.Subsequently, on the same day port was removed.The tubing on the end of the catheter was crimped in the area where it had snapped.Therefore, the investigation is confirmed for the reported fracture, material separation, migration and deformation issues.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B1, b3, b5, g3, h1, h6 (patient, device, method, result).Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|