Product analysis: evaluation determined that a bearing ball detached inside the collet shaft.The likely cause of the failure could not be determined.It was also noted that collet housing is stuck, motor bearings are worn, attachment can't be inserted into the motor and foreign debris in attachment.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|