• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL FUSLITEGU_OLYM-ACMI_3.5MMX3.0M; LIGHT, SURGICAL, FIBEROPTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL FUSLITEGU_OLYM-ACMI_3.5MMX3.0M; LIGHT, SURGICAL, FIBEROPTIC Back to Search Results
Catalog Number 242386
Medical Device Problem Codes Break (1069); Display or Visual Feedback Problem (1184); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: device history review a manufacturing record evaluation was performed for the finished device number, and no non-conformances were identified.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found that device is in normal use condition, however, on the male end, a piece of the optical fiber is detached.The detached piece was not returned.The device was manufactured on 2022 which is more than 2 years old.A functional test was unable to be performed due to the post-manufacturing damage.The observed condition was identified as an end-of-life indicator, damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the fuslitegu_olym-acmi_3.5mmx3.0m would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to end of life; as per ifu, light cables are delicate medical instruments and must be used and handled with care.Process light cables as soon as is reasonably practical following use.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Event or Problem Description
It was reported that during pre-surgery for an unknown procedure, it was observed that the light guide olym-acmi_3.5mmx3.0m device had bad light cord, was very dull and could not get image quality to work.Another like device was used to complete the procedure.There was no delay in the procedure reported.There were no adverse patient consequences reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSLITEGU_OLYM-ACMI_3.5MMX3.0M
Common Device Name
LIGHT, SURGICAL, FIBEROPTIC
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 
3035526892
MDR Report Key21342415
Report Number1221934-2025-00512
Device Sequence Number8959566
Product Code FST
UDI-Device Identifier10886705028818
UDI-Public(01)10886705028818(10)WO187481
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K901035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number242386
Device Lot NumberWO187481
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2025
Initial Report FDA Received Date02/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/25/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-