| Catalog Number |
122128051 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
01/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was revised due to infection and the surgeon did a single stage revision.Surgeon removed all implants, did a thorough i&d, and put in new implants.There was no delay in procedure.Affected side: right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A manufacturing record evaluation (nc search) was performed for the finished device product code: 122128051, lot -3978605 and no non-conformances / manufacturing irregularities were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device product code: 122128051, lot -3978605 and no non-conformances / manufacturing irregularities were identified.
|
| |
|
Search Alerts/Recalls
|