MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. CONTOUR CURVED CUTTER STAPLER; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. CONTOUR CURVED CUTTER STAPLER; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number CS40G
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent 2/10/2025.D4 batch #436c35.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the cs40g device was received with no apparent damage and with a reload loaded in the device and a second reload (b) present.Both reloads (a and b) were received fully loaded with staples, with the washer uncut, with the drivers, and with the knife recess below the reload deck.The ledge of the lockout showed indentation.It is possible that an attempt was made to close the device with a reload not properly loaded or with a spend reload, this condition would lock the device giving the impression that the device will not close.The contour¿ curved cutter is designed with a feature that prevents closure if a used reload, no reload, or an incorrectly inserted reload is in the instrument.The used or misloaded reload module should then be replaced with a new reload or, if no reload was present, a reload should be loaded in the instrument.After following the above steps, any device which does not close either partially or completely should not be used.Please reference the instruction for use for more information.The reload was not properly loaded in the device, as the retaining pin was not connected to the coupler and pushrod.These facts indicate that the reload was removed and reinserted incorrectly and/or incompletely after the retainer was removed.The returned reload was pulled out of the device and placed back on; the device opened and closed without any difficulties.The device was tested for functionality with a returned reload and the device fired, forming all staples and cutting as intended.The cut line and the staple line were completed and the staples meet the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.It should be noted that if the reload is removed from the device, whether the reload is spent or not, and if the staple retainer has already been removed from the reload, the reload cannot be reloaded into the device.Please reference the instructions for use for the complete guide loading and reloading the instrument.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 436c35, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device gcs40g lot number a9d782 and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: 1.Could you please provide more for "gcs40g is cancelled"? 2.Please provide more details for each device malfunction.3.Did the device cut? 4.If the device cut/fired, was the cut line complete? 5.Did the device staple? 6.If the device stapled/fired, was the staple line complete? 7.If the device stapled, was the shape of the staples (b-formation, irregular, unformed)? 8.Did the device close? 9.Did the device fire? 10.Did the device open? 11.Was the device difficult to close on the tissue? 12.Was the device difficult to fire? 13.Was the device difficult to open? 14.On which firing did this occur? 15.Did the tissue retaining pin lock into the correct position? 16.Regarding "patient status/ outcome / consequences: yes,", please provide more details.
 
Event or Problem Description
It was reported that during an unknown procedure the device is not working the surgery was delayed 10 minutes.Changed the stapler.The procedure was successfully completed.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR CURVED CUTTER STAPLER
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key21344334
Report Number3005075853-2025-01058
Device Sequence Number15412996
Product Code GDW
UDI-Device Identifier10705036012405
UDI-Public(01)10705036012405(17)280531(10)A9D782
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K040038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCS40G
Device Lot NumberA9D782
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2025
Initial Report FDA Received Date02/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CR40G.
Patient SexUnknown
-
-