• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS MIDAS REX LEGEND; MOTOR, DRILL, ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS MIDAS REX LEGEND; MOTOR, DRILL, ELECTRIC Back to Search Results
Model Number EM100-A
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
Repair request initiated for device with the report of overheating.It was reported that there was no patient involvement.Repair request escalated to a product event based on reason for return and return in less than 90 days from purchase or repair.
 
Additional Manufacturer Narrative
H3: product analysis: evaluation determined overheating and the maximum temperature was measured to be 109.5°f.The likely cause of failure could not be determined.This regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX LEGEND
Common Device Name
MOTOR, DRILL, ELECTRIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key21345907
Report Number1625507-2025-00185
Device Sequence Number9078384
Product Code HBC
UDI-Device Identifier00763000430887
UDI-Public00763000430887
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K012453
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEM100-A
Device Catalogue NumberEM100-A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2022
Initial Report FDA Received Date02/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/06/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-