| Model Number |
3000TFX21MM |
| Medical Device Problem Code |
Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Insufficient Information (4580)
|
| Date of Event |
01/16/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to edwards for evaluation as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported a patient with a 21mm 3000tfx aortic valve implanted nine (9) years, nine (9) months, underwent valve-in-valve procedure due to aortic stenosis.Tavr was successfully performed with a 20mm 9755rsl transcatheter valve.
|
| |
|
Additional Manufacturer Narrative
|
|
Added information to b5, b7, h6.
|
| |
|
Event or Problem Description
|
|
It was reported a patient with a 21mm 3000tfx aortic valve implanted nine (9) years, nine (9) months, underwent valve-in-valve procedure due to aortic stenosis.Patient presented with shortness of breath.Tavr was successfully performed with a 20mm 9755rsl transcatheter valve.Patient was stable at the end of the procedure.Patient was discharged on pod# 1.
|
| |
|
Additional Manufacturer Narrative
|
|
Added information to h6.The device history record (dhr) review was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient-related factors, including hyperlipidemia, diabetes mellitus.
|
| |
|
Search Alerts/Recalls
|