H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
H11: additional manufacturer narrative: updated: b4, d4, g3, g6, h2, h4, h6.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Regurgitation identified post-procedurally has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Regurgitation occurring during device implantation or post-procedurally is most commonly related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing non-conformance.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.A definitive root cause is unable to be determined, however, patient and likely caused or contributed.
|