Model Number EM200
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis: evaluation could not be performed because shaft broken/cracked.It was noted that worn rear bearings and motor corroded.This finding may have contributed to the user's experience.This regulatory report is being submitted due to a retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Repair request initiated for the device with the report of overheating.It was reported that there was no patient involvement.Repair is being escalated to product event due to reason for return and the repair and return in less than 90 days from purchase or repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX LEGEND
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key21353338
Report Number1625507-2025-00210
Device Sequence Number18421044
Product Code HBE
UDI-Device Identifier00643169505537
UDI-Public00643169505537
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K012453
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEM200
Device Catalogue NumberEM200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/29/2022
Initial Report FDA Received Date02/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/26/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-