Model Number G22554
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Continued: e1 country: brazil street (continued): 492 / 504 ¿ mooca phone: (b)(6).Investigation evaluation: a product evaluation was performed only by the pictures provided with this report because the product said to be involved was not provided to cook for evaluation.A photo of the lot number was not provided.Our evaluation of the photos provided confirmed the report.The photos show the distal end of the device through the clear side of the opened pouch.A brown substance is visible within the clear catheter.The anchor has separated from the catheter at the distal end but remains securely attached at the proximal end of the wire.The anchor appears to have both a long segment and a short segment, therefore, no portion has detached.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the photos provided confirmed that the cutting wire securing component has separated from the catheter without detachment of the cutting wire securing component.However, we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Separation of the cutting wire securing component and the catheter can occur if the tip of the sphincterotome is over flexed.The instructions for use caution the user with the following comment: ¿do not over flex or bow the tip beyond 90 degrees, as this may damage or cause the cutting wire to break.¿ other factors that can contribute to separation of the cutting wire securing component and the catheter include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use contain the following precaution: "do not exercise the handle while the device is coiled or the precurved stylet is in place, as this may cause damage to the sphincterotome and render it inoperable." the instructions for use advise the user with the following statement: ¿upon removing the device from the package, uncoil and straighten the sphincterotome.Carefully remove the precurved stylet from the cannulating tip.¿ prior to distribution, all tri-tome pc triple lumen sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment, no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a cook tri-tome pc triple lumen sphincterotome.It was reported that the nurse stated that the doctor introduced the sphincterotome, but when manipulating its curvature, the cutting wire came loose, making it necessary to use another sphincterotome and the procedure could be continued normally.Images of the device were provided showing the cutting wire anchor detached.[subject of report].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-TOME PC TRIPLE LUMEN SPHINCTEROTOME
Common Device Name
UNIT, ELECTROSURGICAL ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston salem NC
Manufacturer (Section G)
WILSON-COOK MEDICAL INC (COOK ENDOSCOPY)
4900 bethania station rd
winston salem NC
Manufacturer Contact
sabrina o¿brien
4900 bethania station rd
winston salem, NC 
3367440157
MDR Report Key21354938
Report Number1037905-2025-00114
Device Sequence Number18101835
Product Code KNS
UDI-Device Identifier00827002225541
UDI-Public(01)00827002225541(17)270520(10)W4838748
Combination Product (Y/N)N
PMA/510(K) Number
K172665
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG22554
Device Catalogue NumberTRI-25
Device Lot NumberW4838748
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/15/2025
Initial Report FDA Received Date02/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/20/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENDOSCOPE, UNKNOWN MAKE OR MODEL
Patient Age41 YR
Patient SexMale
-
-