Catalog Number 339.31
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that the ophthalmic probe had a problem with the electrocoagulation head, there was no response when stepping during surgery.The procedure was completed on the same day after replacing it with a new diathermy probe.There were no abnormalities in the patient.
 
Additional Manufacturer Narrative
Additional information provided in sections d.9., h.3., h.6., and h.11.A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A review of the database for the non-conformance events and the complaints showed that the given lot number was not concerned by any deviation and that there are no additional complaints associated with it.The investigation is completed based on the evaluation of the returned diathermy probe documented below.The returned instrument was received in the pouch with the sticker label showing the batch information and with a plastic tubing put on the tip for protection.The instrument was visually inspected and showed no evident abnormality.The electrical resistance measurements indicate that there was insufficient contact, leading to the instrument not functioning.When the instrument was opened to visually inspect the inner contacts, some of them were unintentionally destroyed (destructive instrument testing).No other defects or deficiencies could be identified on the contacts after opening the instrument.As the pouch was opened by the customer, the product was handled.Therefore, the point in time when the malfunction occurred can no longer be determined, nor can the strain put on the device be reconstructed.The customer¿s complaint is confirmed, but a root cause for the reported issue cannot be determined.The root cause cannot be identified conclusively, as the tyvek pouch was opened by the customer, the product was handled.The chain of events that led to the reported malfunction can therefore no longer be determined.The exact root cause for the customer's reported event is unknown, therefore, specific action cannot be taken.This complaint has been reviewed and future data will be monitored for evidence of adverse trending and further action will be taken, as appropriate.At a minimum, this will include completing reviews of complaint class report levels on a monthly basis.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIATHERMY PROBE DSP
Common Device Name
UNIT, PHACOFRAGMENTATION, ACCESSORY, PROBE
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21355056
Report Number3003398873-2025-00184
Device Sequence Number18393086
Product Code HQC
UDI-Device Identifier07612717071384
UDI-Public(01)07612717071384(17)261130(11)231219(10)16E3CN
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K063583
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number339.31
Device Lot Number16E3CN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/24/2025
Supplement Date Received by Manufacturer03/10/2025
Initial Report FDA Received Date02/11/2025
Supplement Report FDA Received Date04/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/19/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-