On 24-jan-2025, abbott point of care was contacted by a customer regarding i-stat troponin cartridges that yielded descrepant false positive result of 0.24 ng/ml on a patient.There was no patient information, nor details surrounding the event.Return product is available for investigation.Method: date: tested: result: i-stat, (b)(6) 2025, 12:36, 0.28 ng/ml.Vitros 7600, ni.Ni, <0.012 ng/ml.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggested that product was not performing within the variability.There is no evidence the pateint suffered an mi.The investigation is underway.Per i-stat system manual: art: 715595-00v, reportable range: 0.00 - 50.00, reference range: 0.00 - 0.03 (represents the 0 to 97.5% range of results), reference range: 0.00 - 0.08 (represents the 0 to 99% range of results).
|