| Model Number |
434893 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A pk papyrus covered stent system was selected for treatment of a perforation in the mid lad.The stent dislodged from the delivery system and was retrieved.It was not possible to deliver an additional stent.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected pk papyrus stent system was not returned to biotronik and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form) and the production documentation were reviewed to establish whether a device deficiency contributed to this event.Review of the production documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the provided documentation and the conducted review, no manufacturing related root cause could be identified.The treating physician rated the outcome as both no device- and no procedure-related complication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
|
| |
|
Event or Problem Description
|
|
A pk papyrus covered stent system was selected for treatment of a perforation in the mid lad.The stent dislodged from the delivery system and was retrieved.It was not possible to deliver an additional stent.
|
| |
|
Search Alerts/Recalls
|