MSD BARTLETT MFG CD HORIZON X10 CROSSLINK; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 8110530
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/30/2024
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from a user facility (uf) via manufacturer representative regarding spinal product used in unknown spinal therapy.It was reported that the device was broken.There is no patient involved in the event and no further complications or symptoms were reported.
 
Additional Manufacturer Narrative
E: first name and last name of initial reporter is unknown.H3: product analysis of part# 8110530, lot# id07d010 visual inspection confirmed a portion of the hex driver has broken.Optical inspection revealed a flat brittle surface fracture.This type of damage is consistent with bend stress overload.H11: this regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON X10 CROSSLINK
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MSD BARTLETT MFG
2975 brother blvd
bartlett TN 38133
Manufacturer (Section G)
MSD BARTLETT MFG
2975 brother blvd
bartlett TN 38133
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key21355653
Report Number1030489-2025-00716
Device Sequence Number9003558
Product Code LXH
UDI-Device Identifier00721902744600
UDI-Public00721902744600
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number8110530
Device Catalogue Number8110530
Device Lot NumberID07D010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/30/2024
Initial Report FDA Received Date02/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-