• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. FW BPB #5 SUTR,BLU W/NDL; POLYETHYLENE SYNTHETIC SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. FW BPB #5 SUTR,BLU W/NDL; POLYETHYLENE SYNTHETIC SUTURE Back to Search Results
Model Number FW BPB #5 SUTR,BLU W/NDL
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On (b)(6) 2025, it was reported by a sales representative via (b)(4) that (qty.2) of an ar-7211 fiberwire #5 braided polyblend suture broke during use.No pieces broke off in the patient.The case was completed successfully.This was discovered during a quad repair procedure on (b)(6) 2025, with no reported patient harm.Additional information has been requested.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.The complaint device was not received for evaluation.Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude the most likely cause.The most likely cause of the reported failure is user error, specifically damage caused by crushing or crimping due to the use of surgical instruments such as forceps or needle holders.The complaint allegation was confirmed based on the customer's attached image, which shows a blue suture with frayed tips, a missing curved needle, and filaments protruding from the suture strand.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FW BPB #5 SUTR,BLU W/NDL
Common Device Name
POLYETHYLENE SYNTHETIC SUTURE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21356475
Report Number1220246-2025-00380
Device Sequence Number8959834
Product Code GAT
UDI-Device Identifier00888867041417
UDI-Public(01)00888867041417(11)230901(17)280831(10)29460
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K122374
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFW BPB #5 SUTR,BLU W/NDL
Device Catalogue NumberAR-7211
Device Lot Number29460
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2025
Supplement Date Received by Manufacturer09/03/2025
Initial Report FDA Received Date02/11/2025
Supplement Report FDA Received Date09/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-