| Model Number |
K063 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that review was requested of the data from this pacemaker system.Technical services (ts) analyzed device episode data and found many ventricular episodes due to oversensing of myopotential noise on the right ventricular (rv) lead channel for the past four years.No additional adverse patient effects were reported.Currently, this pacemaker system remains in service.
|
| |
|
Event or Problem Description
|
|
It was reported that review was requested of the data from this pacemaker system.Technical services (ts) analyzed device episode data and found many ventricular episodes due to oversensing of myopotential noise on the right ventricular (rv) lead channel for the past four years.No additional adverse patient effects were reported.Currently, this pacemaker system remains in service.According to additional information, there was an alert for out of range shock lead impedance values.Technical services (ts) noted that the values were less than 200 ohms.No additional adverse patient effects were reported.Currently, this pacemaker system remains in service.
|
| |
|