| Model Number |
99331 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, the attached y adapter disconnected during patient use which caused a leak.It is unknown how long it was disconnected; there was no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The product involved in the report was discarded and therefore is not available for return.A review of the device history record is in-progress.All information reasonably known as of 11 feb 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
It was reported, the attached y adapter disconnected during patient use which caused a leak.It is unknown how long it was disconnected; there was no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d1; g1.Investigation results: d4; h5; h6.Additional information: na.The product involved in the report was discarded and therefore is not available for return; however, the reporter provided photographic evidence for evaluation.Review of the photographic evidence confirmed the issue as reported; the root cause was traced to the manufacturing process.Processes are in place to prevent the event from reoccurring.The device history record for lot 30327235 was reviewed and the product was produced according to product specifications.All information reasonably known as of 20 jun 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
|
| |
|
Search Alerts/Recalls
|