| Model Number |
406114 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a procedure, after insertion in the patient, the side port disconnected from the sheath.There were no adverse patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h3, h6, h11.One 8f fast-cath introducer sheath and dilator were received for evaluation.The reported event of sideport tubing detachment was confirmed.The extension tubing had detached from the sideport of the hemostasis body.A corrective action was initiated to investigate the failure mode of the sideport detachments.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, e1, h6.
|
| |
|
Event or Problem Description
|
|
During a procedure, after insertion in the patient, the side port disconnected from the sheath causing minor bleeding with no intervention required.There were no adverse patient consequences.
|
| |
|