| Model Number |
24970A |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: performance data collected from the mobile programmer was received and analyzed.Analysis of the data/database found the customer comment of the mobile programmer displayed noise on the atrial lead was confirmed.It was also determined that the telemetry connection was lost.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the mobile programmer displayed noise on the atrial lead.Troubleshooting steps were advised.No patient complications have been reported as a result of this event.Upon analysis it was noted that there was a loss of telemetry connection.
|
| |
|