| Model Number |
8670020 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis: part# 8670020; lot# (b)(6) visual and optical examination identified it appears that part of the driver's tip has been sheared off and the rest of the tip is twisted.The metal deformation gives indication that the failure was the result of torsional overload.This regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having olif l3-5 for d egenerative disc disease l3-5.It was reported that while the surgeon was removing previously placed hardware at l5, the screwdriver broke.A new driver was opened and the case carried on with no issue.There was no patient symptom reported.There were no further complications reported regarding the event.Additional information was received from the manufacturer representative that the device came in contact with the patient.
|
| |
|
Search Alerts/Recalls
|