| Model Number |
SN60WF |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported with a description of when loading the intraocular lens into the injector the surgeon noticed that the apex of the lens had cracks visible under a microscope.The procedure was completed with another lens.There was no patient contact.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in d.9., h.3., h.6.And h.11.The lens was returned for evaluation.Solution is dried on the lens.One haptic is broken in the gusset area and returned adhered to the optic by solution.The opposite haptic is broken in the distal tip and folded under the optic portion.A small crack is observed on the optic near the edge.Product history records were reviewed and the documentation indicated the product met release criteria.Associated products were not provided.It is unknown if qualified products were used.The reported lens damage was observed.All lenses are 100percentage inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met companys release criteria.Based on our observation, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.Associated products were not provided.It is unknown if qualified products were used.Company foldable iols are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The instructions for use (ifu) instructs using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd filled cartridge.The lens should be inserted until the optic is a little more than half way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.Due diligence has been performed in an attempt to obtain further information on this event.The reporter has not responded to request for follow up information.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|