• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SN60WF
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported with a description of when loading the intraocular lens into the injector the surgeon noticed that the apex of the lens had cracks visible under a microscope.The procedure was completed with another lens.There was no patient contact.Additional information has been requested.
 
Additional Manufacturer Narrative
Additional information was provided in d.9., h.3., h.6.And h.11.The lens was returned for evaluation.Solution is dried on the lens.One haptic is broken in the gusset area and returned adhered to the optic by solution.The opposite haptic is broken in the distal tip and folded under the optic portion.A small crack is observed on the optic near the edge.Product history records were reviewed and the documentation indicated the product met release criteria.Associated products were not provided.It is unknown if qualified products were used.The reported lens damage was observed.All lenses are 100percentage inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met companys release criteria.Based on our observation, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.Associated products were not provided.It is unknown if qualified products were used.Company foldable iols are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The instructions for use (ifu) instructs using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd filled cartridge.The lens should be inserted until the optic is a little more than half way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.Due diligence has been performed in an attempt to obtain further information on this event.The reporter has not responded to request for follow up information.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
++++++
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21363839
Report Number1119421-2025-00365
Device Sequence Number8465322
Product Code HQL
UDI-Device Identifier00380655093221
UDI-Public(01)00380655093221(17)281029(11)231101(10)15740428
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.215
Device Lot Number15740428
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/17/2025
Supplement Date Received by Manufacturer03/10/2025
Initial Report FDA Received Date02/12/2025
Supplement Report FDA Received Date04/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-