|
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.One unpackaged, ar-611-4, ibalance® uka sportsplasty¿, tibial impactor, batch number: 051752, was received for investigation.Visual evaluation of the returned device noted that fragments from the distal end had broken off.Additionally, the impactor head had significant damage: strike marks and discoloration.Functional testing was not performed due to the damage to the device.The most likely cause for the reported failure can be attributed to overstressing a device that is damaged naturally and inevitably as a result of normal wear or aging from repeated usage/reprocessing.Manufactured date: 15-feb-2018.
|