| Catalog Number |
100/897/090CZ |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device leaked when inflated.The date of the event was (b)(6) 2024
at 10:30.The event occurred during use on patient for less than 10 minutes.There was no patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1 (b)(6).G4: 510k is unknown.No product information has been provided.Device has not been returned to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for investigation in used condition with no visible signs of damage.Functional testing was completed where the cuff was inflated with air and submerged into a water bath.A leak was observed between the balloon and the inlet tube confirming the customer complaint.Root cause was not confirmed but the probable cause was attributed to a manufacturing issue.A review of the device history records show that there were no anomalies or non conformities during the manufacturing process. one device was received for evaluation.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.Due to this, no root cause was determined as the device worked as expected according to the instructions for use.Device history review of the reported lot number found no anomalies or discrepancies during the manufacturing process.
|
| |
|
Event or Problem Description
|
|
It was reported that the device leaked when inflated.The date of the event was 20-jan-2024 at 10:30.The event occurred during use on patient for less than 10 minutes.There was no patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint was incorrectly filed under this registration number, please reference mrn: 3012307300-2025-05698 for details pertinent to this event.
|
| |
|