|
Section b2: outcomes attributed to adverse corrected.Section d9: a portion of the percutaneous lead returned for evaluation.Section g2: report source corrected.Manufacturer¿s investigation conclusion: incidental findings: damage to the driveline outer jacket was observed approximately ~3, 5, 13, and 15-19" from the controller connector.Evaluation of the patient¿s replaced portion of the driveline from heartmate ii left ventricular assist system (lvas), serial number (b)(6), confirmed wire damage which could have contributed to the low speed and pump stop events observed within the submitted log file.A distal end driveline replacement was performed on (b)(6) 2025.Approximately 21¿ of the external portion of the driveline was returned for evaluation.The pins of the controller connector appeared unremarkable.The driveline was tested for electrical continuity in the condition that it was received, and all wires were found to be electrically intact.Following careful removal of all layers, microscopic examination of the underlying wires revealed multiple breaches in the insulation of the orange wire approximately 1" from the controller connector.One of the breaches in this location appeared consistent with an area of shorting.Additional breaches in the insulation of the brown, yellow, and green wires were also observed approximately 1" from the controller connector; however, no obvious evidence of shorting was observed in these locations.The remaining wires, as well as the remaining portions of the damaged wires, were submerged in a saline bath solution for high potential (hi-pot) testing to further verify the integrity of each wire's insulation.This test did not reveal any additional areas of current leakage.Although a specific cause could not conclusively be determined, the observed wire breaches appeared to be the result of repetitive flexing of the driveline in an area approximately 1¿ from the controller connector.If any of the exposed conductors made contact with the metal braided shield while operating on the power module, a short-to-shield would have resulted, contributing to the low speed and pump stop events observed within the submitted log file.Additionally, if any of the exposed conductors from wires of different motor phases made contact with each other, or simultaneous contact with the metal braided shield, a phase-to-phase short would have resulted, which could have also contributed to the pump behavior observed within the submitted log file.The relevant sections of the device history records for (b)(6) and the driveline, serial number (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.The current revision of the instructions for use (ifu) can be found on the eifu page of the abbott website.The heartmate ii left ventricular assist system (lvas) instructions for use (ifu), rev.C, is currently available.Section 6, ¿patient care and management¿, discusses wear and tear to the driveline, contains information on ¿caring for the driveline¿, and provides possible indications of driveline damage as well as how to respond to such events.Section 5, "surgical procedures", cautions the user that sharp bends, twists, or kinks in the driveline may make it more susceptible to wear and fatigue over time.Section 8, ¿equipment storage and care¿, also contains information on ¿care of the driveline,¿ and provides possible indications of driveline damage.Section 7, ¿alarms and troubleshooting¿, addresses all system controller alarms (including pump off and low speed alarms) and how to respond to such events.The heartmate ii lvas patient handbook, rev.C, is also currently available.The patient handbook contains a section on ¿caring for the driveline¿; however, all heartmate ii left ventricular assist device (lvad) drivelines have the potential for wire/shield breakdown to occur dependent upon length of use and patient handling.Additionally, the patient handbook outlines proper driveline maintenance.If driveline damage is suspected, the user is instructed to call their hospital contact immediately.The handbook also addresses all system controller alarms (including pump off and low speed alarms) and how to respond to such events.No further information was provided.The manufacturer is closing the file on this event.
|