• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACCLARENT, INC. RELIEVA SPINPLUS NAV; INSTRUMENT, ENT MANUAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACCLARENT, INC. RELIEVA SPINPLUS NAV; INSTRUMENT, ENT MANUAL SURGICAL Back to Search Results
Catalog Number RSP0616MFSN
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Cerebrospinal Fluid Leakage (1772)
Date of Event 01/21/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a trudi¿ navigation case in the operating room (or), a cerebrospinal fluid or csf leak was noticed in the patient.The caller reported that there were no issues or errors observed on the trudi¿ navigation system.The caller reported that during the procedure, the physician had noticed a small cerebrospinal fluid leak in the patient.The caller reported that the medical intervention provided to the patient was a mucosal patch or graft.The caller reported that the patient was then sent to the hospital for further monitoring.The caller was unable to confirm if the patient is in stable condition at this time.It was also reported that the following acclarent products were used during this case: relieva spinplus® nav balloon - rsp0616mfsn, trudi¿ probe (70 degree) - tdp0705, trudi¿ nav suction device (0 degree) - tdns000z, trudi¿ nav suction device (70 degree) - tdns070z, trudi¿ nav suction device (90 degree) - tdns090z.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, weight, and ethnicity were not provided.D.2b: procode is lrc/pgw.Based on complaint information, the device is not available to be returned for analysis.Product analysis cannot be conducted as the product was not returned for analysis.A manufacturing record evaluation was performed and no non-conformances related to the complaint was found during the review.At this time, the provided available information is nonspecific and very limited.Multiple attempts have been made to obtain additional details regarding this event.However, no additional information has been made available.Since this adverse event required medical or surgical intervention to preclude permanent impairment of a body function or permanent damage to a body structure, it is considered serious and reportable.Should additional information become available, this case will be re-assessed and notes will be updated.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by acclarent, or its employees that the report constitutes an admission that the product, acclarent, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIEVA SPINPLUS NAV
Common Device Name
INSTRUMENT, ENT MANUAL SURGICAL
Manufacturer (Section D)
ACCLARENT, INC.
31 technology drive
irvine CA 92618
Manufacturer (Section G)
ACCLARENT, INC.
31 technology drive
irvine CA 92618
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key21367347
Report Number3005172759-2025-00014
Device Sequence Number8481812
Product Code LRC
UDI-Device Identifier10705031241015
UDI-Public(01)10705031241015(17)260527(10)240527C-PC
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K171687
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/27/2026
Device Catalogue NumberRSP0616MFSN
Device Lot Number240527C-PC
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2025
Initial Report FDA Received Date02/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PATIENT TRACKER DISPOSABLE 1PK; TRUDI NAVIGATION SYSTEM
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient SexUnknown
-
-