|
Visual inspection: during the investigation, no significant deformations or damage of the valve were determined.Permeability test: a permeability test has shown that all components are permeable.Wählen sie ein element aus.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the progav 2.0 does not operate within the accepted tolerance in the horizontal position.The shuntassistant operates within the specified tolerances in the vertical position.An accelerated outflow of progav 2.0 could be determined.Adjustment test: the progav 2.0 was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected; however, the breaking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valves, deposits were found in both valves.Results: based on our investigation results, we can determine an accelerated outflow and a non-adjustability at the progav 2.0.The visible deposits in the valve have led to the functional impairment.Furthermore, we can determine visible deposits in the shuntassistant.These deposits had no effect upon the specifications at the time of the examination.Deposits caused by substances naturally present in the patient's body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic material can affect the integrity of the valve.The examination of the return has been completed with the drafting of this report.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|