AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS Back to Search Results
Model Number DIB00
Medical Device Problem Codes Break (1069); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/16/2024
Type of Reportable Event Malfunction
Event or Problem Description
During implantation of the pre-loaded dib00 model intraocular lens (iol) the surgical technician experienced a loading issue and the haptic was broken; the lens was partially inserted.There was no incision enlargement, no medical attention, no procedural delays, no sutures, and no vitrectomy during the initial procedure.It is unknown if medication was prescribed and unknown if another iol was implanted.Patient's daily activities were not significantly affected.Patient status post procedure was reported as fully recovered.The iol directions for use were followed.Patient information cannot be provided due to personal data privacy legislation/policy.No further information is available.
 
Additional Manufacturer Narrative
Additional information: section a-2 patient age/date of birth: information cannot be provided due to personal data privacy legislation/policy.Section a-3a patient sex: information cannot be provided due to personal data privacy legislation/policy.Section a-3b patient gender: information cannot be provided due to personal data privacy legislation/policy.Section a-4 patient weight: information cannot be provided due to personal data privacy legislation/policy.Section a-5 patient ethnicity: information cannot be provided due to personal data privacy legislation/policy.Section a-6 patient race: information cannot be provided due to personal data privacy legislation/policy.Section d-6a date implanted: not applicable as the iol was not implanted.Section d-6b date explanted: not applicable as the iol was not implanted; therefore, not explanted.Section h-3 device evaluated by manufacturer: the complaint device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch report will be filed.An attempt was made to contact the customer account requesting additional information regarding complaint however, to date no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d-9: device available for evaluation: yes.Section d-9: device date returned to manufacturer: 15-may-2025.Section h-3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received loose on a piece of medical foam.No handpiece was received for evaluation.During a visual inspection, the device was inspected under magnification.The complaint lens was received coated in viscoelastic residue and cut in half with one lens half missing.The lens was cleaned revealing the lens was scratched; one haptic was observed to be chipped.Based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Conclusion: based on the complaint investigation results, the product was released within specifications.No product deficiency or product malfunction was identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key21370695
Report Number3012236936-2025-000042
Device Sequence Number15573897
Product Code HQL
UDI-Device Identifier05050474731721
UDI-Public(01)05050474731721(17)251117
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/17/2025
Device Model NumberDIB00
Device Catalogue NumberDIB00U0185
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2025
Supplement Date Received by Manufacturer06/20/2025
Initial Report FDA Received Date02/12/2025
Supplement Report FDA Received Date07/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-