| Model Number |
DIB00 |
| Medical Device Problem Codes |
Break (1069); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During implantation of the pre-loaded dib00 model intraocular lens (iol) the surgical technician experienced a loading issue and the haptic was broken; the lens was partially inserted.There was no incision enlargement, no medical attention, no procedural delays, no sutures, and no vitrectomy during the initial procedure.It is unknown if medication was prescribed and unknown if another iol was implanted.Patient's daily activities were not significantly affected.Patient status post procedure was reported as fully recovered.The iol directions for use were followed.Patient information cannot be provided due to personal data privacy legislation/policy.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section a-2 patient age/date of birth: information cannot be provided due to personal data privacy legislation/policy.Section a-3a patient sex: information cannot be provided due to personal data privacy legislation/policy.Section a-3b patient gender: information cannot be provided due to personal data privacy legislation/policy.Section a-4 patient weight: information cannot be provided due to personal data privacy legislation/policy.Section a-5 patient ethnicity: information cannot be provided due to personal data privacy legislation/policy.Section a-6 patient race: information cannot be provided due to personal data privacy legislation/policy.Section d-6a date implanted: not applicable as the iol was not implanted.Section d-6b date explanted: not applicable as the iol was not implanted; therefore, not explanted.Section h-3 device evaluated by manufacturer: the complaint device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch report will be filed.An attempt was made to contact the customer account requesting additional information regarding complaint however, to date no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d-9: device available for evaluation: yes.Section d-9: device date returned to manufacturer: 15-may-2025.Section h-3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received loose on a piece of medical foam.No handpiece was received for evaluation.During a visual inspection, the device was inspected under magnification.The complaint lens was received coated in viscoelastic residue and cut in half with one lens half missing.The lens was cleaned revealing the lens was scratched; one haptic was observed to be chipped.Based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Conclusion: based on the complaint investigation results, the product was released within specifications.No product deficiency or product malfunction was identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|