Additional information was received, and section h 11 should be reported as: the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap/ bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles.No other clinical information or medical interventions were reported.The manufacturer's investigation is ongoing.The device was returned to the third-party service center for further evaluation.During the evaluation of the device at the third-party service centre, the device was visually inspected and found no evidence of foam degradation.During the evaluation, secondary findings was not observed.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the third-party service centre.There was no error code found.The third-party service center concludes that they couldn't confirm the customer's allegation and there were no visible foam degradation and unit got scrapped.Evaluation method code, evaluation results code, component code and conclusion code grid has been updated.
|