• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR BONE ANCHORED SOLUTIONS AB BA300 ABUTMENT 12MM; COCHLEAR BAHA CONNECT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR BONE ANCHORED SOLUTIONS AB BA300 ABUTMENT 12MM; COCHLEAR BAHA CONNECT SYSTEM Back to Search Results
Model Number 93357
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Skin Infection (4544); Implant Pain (4561)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Transient pain is a known medical complication where most cases can be resolved with clinical case management.Known reasons for cause of pain are abnormal skin sensitivity, earlier trauma to the implanted area, lack of adequate bone quality or other generalized diseases.It might also be due to a loose implant magnet, causing pain around the implanted area when using the baha sound processor.Adverse skin reactions around the baha abutment, ranging from slight redness to infected tissue, are common complications with baha implants, and can occur at any time following implantation.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.
 
Event or Problem Description
Per the clinic, the patient experienced pain and an infection at the abutment site (specific date not reported).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BA300 ABUTMENT 12MM
Common Device Name
COCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke 435 3 3
SW   435 33
Manufacturer Contact
ashwinii chelvan
unit ug-1, vertical podium
no. 8 jalan kerinchi
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key21373127
Report Number6000034-2025-00476
Device Sequence Number8550743
Product Code LXB
UDI-Device Identifier09321502033435
UDI-Public(01)09321502033435(10)COH1511866(17)270110
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K100360
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number93357
Device Catalogue Number93357
Device Lot NumberCOH1511866
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/22/2025
Initial Report FDA Received Date02/13/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-