| Model Number |
AH760HS |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The patient alleges visualization of particles in device.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer¿s investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The patient alleges visualization of particles in device.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.Evaluation method code, evaluation results code and conclusion code grid has been updated.
|
| |
|
Additional Manufacturer Narrative
|
|
A bipap auto biflex device was returned to a third-party service center for evaluation.During the evaluation of the device at the third-party service center, the device was visually inspected, and evidence of foam particles was found inside the blower kit.Additionally, power connecter was destroyed and the evaluation of the device found error codes.Lastly, the device was scrapped by the service center after the evaluation was completed.In this report box h has been updated/ corrected.The affected products with field complaints alleging pe-pur foam degradation have undergone the establishment of complaint codes specifically identifying foam degradation, for system one.A field correction action(2021-06-a) was initiated, and medical device recall letters were issued.A field action was initiated to replace pe-pur foam in affected products; degradation was added to the respective dfmeas and the risk management file ¿ 2203077 v27 (systemone) was updated to include hazards associated with foam degradation based on complaints analyzed through various health hazard evaluations (hhes).Risk tags ¿ degradation01, toxic02 and particulate01 reflect the safety risk of design containing the pe-pur foam.These complaints were monitored and trended in accordance with the complaint trending procedures.If the complaints were determined to meet the criteria for escalation, they were escalated for risk assessment through the post market clinical escalation process.As of the date of submission for this report, there is no indication that complaints for the failure in question have led to unacceptable levels of severity or probabilities of harm, and therefore no further action will be pursued.".
|
| |
|
Search Alerts/Recalls
|
|
|