• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG PROTEGO SD 65/18; ICD LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG PROTEGO SD 65/18; ICD LEAD Back to Search Results
Model Number 399410
Medical Device Problem Codes Fracture (1260); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 02/10/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
This rv lead was capped due to fracture resulting in multiple shocks.Should additional information be received, this file will be updated.
 
Additional Manufacturer Narrative
Combination product: yes.The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Additional Manufacturer Narrative
Combination product: yes.Upon receipt, the lead under complaint was found cut 21 cm distal to the df4 connector pin.The proximal fragment was received for analysis.The distal fragment was not returned for analysis.It is reasonable to assume that the lead was cut during the surgery.The returned lead fragment was analyzed.The inspection revealed that the insulation was, apart from the present cut, free of breaches.Further analysis of the returned lead fragment did not reveal any irregularity that might have contributed to the reported clinical observation.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
 
Event or Problem Description
This rv lead was capped due to fracture resulting in multiple shocks.Should additional information be received, this file will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGO SD 65/18
Common Device Name
ICD LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
DE  12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key21374322
Report Number1028232-2025-00764
Device Sequence Number8623618
Product Code NVY
UDI-Device Identifier04035479136116
UDI-Public(01)04035479136116(17)180430
Combination Product (Y/N)Y
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P980023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2018
Device Model Number399410
Device Catalogue Number399410
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/11/2025
Supplement Date Received by Manufacturer02/14/2025
05/20/2025
Initial Report FDA Received Date02/13/2025
Supplement Report FDA Received Date02/14/2025
05/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/02/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age84 YR
Patient SexFemale
-
-