• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI CAPSUREFIX LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI CAPSUREFIX LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 5568-53
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 12/01/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the implantable pulse generator (ipg), right atrial (ra) lead and right ventricular (rv) lead were explanted four weeks post implant due to an infection.The patient was later implanted with a leadless pacemaker.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX LEAD
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key21374442
Report Number2649622-2025-04247
Device Sequence Number3848847
Product Code DTB
UDI-Device Identifier00613994776853
UDI-Public00613994776853
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P930039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date12/16/2017
Device Model Number5568-53
Device Catalogue Number5568-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2025
Initial Report FDA Received Date02/13/2025
Date Device Manufactured01/13/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ADDRL1 IPG, 407458 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age71 YR
Patient SexMale
-
-